Patient Monitor CMS9200 & Pr/Nibp/Spo2/Temp/ECG
R20,683.99
CMS9200 features in high cost-performance, high safety and low power consumption. To ensure monitoring quality and reduce operation risk, it adopts full-isolation(floating), ECG defibrillation-proof protection, ECG anti-high frequency surgical unit and NIBP dual-overpressure protection
Description
Patient Monitor CMS9200 & Pr/Nibp/Spo2/Temp/ECG
FEATURES:
- Display mode: 15” color TFT LCD with high-resolution.
- Be applicable for adult, pediatric and neonatal for all-round monitoring.
- Operation interface with Chinese and English.
- Waveform and data can be set optionally.
- Storage of 72-hour trend data, and review of 40-second holographic waveform.
- Full-lead ECG display.
- 7-lead ECG displays in a screen, and ECG waveforms series display.
- Anti-high frequency surgical unit, defibrillation-proof(requirement for special leads).
- Adopt digital SpO2 technology, which has strong anti-interference and anti-weak filling capability.
- Function of NIBP review, storage for up to 400 NIBP data.
- Network: connecting with central station, other Bed observation and software updating.
- Network connection mode: wireless and wired.
- Built-in rechargeable battery for uninterrupted monitoring.
- Standard parameters of ECG, RESP, NIBP, SpO2, TEMP and PR.
- Built-in printer(optional).
- Standard parameters of ECG, RESP, NIBP, SpO2, and TEMP.
- IBP, CO2 and thermal printer are optional.
PERFORMANCE:
- ECG
Lead Mode 3-lead and 5-lead are optional
Lead Selection I, II, III, avR, avL, avF, V
Wave 5-lead: 2 channels
3-lead: 1channel
Gain ×0.25, ×0.5, ×1, ×2
HR Measuring and Alarm Range
Range 15 ~ 300 bpm
Accuracy ±1% or ±1bpm, which is greater
Alarm Accuracy ±2bpm
Resolution 1 bpm
- ST Segment Monitoring
Measuring and Alarm Range -0.6 mV~ + 0.8 mV
- ARR
ARR Detecting Type ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC
Alarm Available
Review Available
Scan Speed for ECG Waveform 12.5mm/s, 25mm/s, 50mm/s
- Respiration (RESP)
Method R-F(RA-LL) Impedance
Measuring and Alarm Range for Resp. Rate:
Adult 0~120rpm
Pediatric and Neonatal 0~150rpm
Resolution 1 rpm
Measuring Accuracy 2 rpm
Apnea Alarm 10 ~ 40 S
- SpO2
Measuring and Alarm Range 0 ~ 100 %
Resolution 1 %
Accuracy:
70% ~ 100% ±2%
0% ~ 69% unspecified
- PR
Measuring and Alarm Range 0~250bpm
Resolution 1bpm
Measuring Accuracy ±2bpm or ±2%, which is greater
Alarm Accuracy ±2bpm
- NIBP
Method Oscillometry
Mode Manual, Auto, continuous
Measuring Interval in AUTO Mode 1 /2 /3 /4 /5/ 10/ 15/ 30 /60 /90 Min
Measuring Period in Continuous Mode 5 Min
Measuring and Alarm Range 10~270mmHg
Resolution 1 mmHg
Accuracy ±10% or ±8mmHg, which is greater
Over-pressure Protection:
Adult Mode 315±10 mmHg
Pediatric Mode 265±10 mmHg
Neonatal Mode 155±10 mmHg
- TEMP
Channel dual-channel
Measuring and Alarm Range 0 ~ 50C
Resolution 0.1C
Accuracy ±0.1C
Optional
- EtCO2
Method Sidestream or Mainstream Measuring Range for CO2 0~150mmHg
Resolution for CO2:
0.1 mm Hg 0 to 69 mm Hg
0.25 mm Hg 70 to 150 mm Hg
Accuracy for CO2:
0 – 40 mm Hg ±2 mm Hg
41 – 70 mm Hg ±5%
71 – 100 mm Hg ±8%
101 – 150 mm Hg ±10%
Respiration Rate>80BPM ±12%
AwRR Range 2~150 rpm
AwRR Accuracy ±1BPM
Apnea Alarm Available
- IBP
Channel dual-channel
Label ART, PA, CVP, RAP, LAP, ICP, P1, P2
Measuring and Alarm Range -50~350 mm Hg
Resolution 1 mm Hg
Accuracy ±2% or 1mm Hg, which is greater
- Recorder
Print Mode thermal array
Recording Width 48mm
Paper Speed 25mm/s, 50mm/s
Quantity of Printing Waveform dual-channel
- Physical Characteristic:
Dimension 365 x 150 x 340 (mm)
Net Weight 4.0K
Accessories:
- Adult SpO2 probe(5-pin)
- Adult NIBP cuff
- Extending tube for NIBP
- ECG lead
- ECG electrode
- Temperature probe
- IBP sensor(optional)
- IBP integration module(optional)
- IBP integration module(optional)
- Wall Bracket (optional)
- IBP cable(optional)
- CO2 module(optional)
- CO2 collection pipe(optional)
- Power cord
- User Manua
Reviews